• Product Description

    Testo 327-1 is your introductory instrument to flue gas analysis. It measures combustion efficiency, °C, O2, CO2, CO and flue draught. The strong LED background light in the 4 line display guarantees an easy-to-read display even if lighting conditions are unfavourable. The analyser stands out on account of its easy menu navigation and ergonomic housing as well as its durability.

    Features

    Easy menu navigation

    4 line segment display

    LED display light

    Easy IR printout

    Built-in condensate trap

    EN 50379 Part 2 for O2, °C, hPa

    EN 50379 Part 3 for CO

    Small Li-ion rechargeable battery (1200 mA, lifetime of 5 h) can be recharged inside or outside instrument

    Fast probe connection using single probe plug

    Ambient CO measurement using flue gas probe

    Separate AT temperature measurement

    Undiluted CO measurement (can be stored)

    Draught measurement

    6-8 fuels (country-specific (e.g. UK=6))

    IP 40

    What Is Included:

    Recharable Battery

    Calibration Protocol

    Flue Gas Probe (180mm X 6mm diameter)

    Differential Hose Kit

    Differential Temp kit

    Testo Printer

    Carry Case

    Item Specification

    Storage temperature -20 to +50 °C
    Operating temperature -5 to +45 °C
    Battery life > 5 h
    Weight Approx. 500 g
    Dimensions 216 x 68 x 47 mm
    Probe - Type K
    Meas. Range -40 to +600 °C
    Accuracy ±0,5 °C (0.0 to +99.9 °C)
    ±0,5 % of mv (100 to +600 °C)
    Resolution 0.1 °C
    Probe - Draught sensor
    Meas. Range ±40 hPa
    Accuracy ±0.02 hPa (-0.50 to +0.60 hPa)
    ±0.03 hPa (+0.61 to +3.00 hPa)
    ±1.5% of mv (+3.01 to +40 hPa)
    Probe - Exhaust gas loss measurement
    Meas. Range 0 to 99.9%
    Resolution 0.1 %
    Probe - Electrochemical sensor
    Meas. Range 0 to 4000 ppm
    Accuracy ±20 ppm (0 to 400 ppm)
    ±5% of mv (401 to 1000 ppm)
    ±10% of mv (1001 to 4000 ppm)
    Resolution 1 ppm
    Probe - Electrochemical sensor
    Meas. Range 0 to 21 Vol. %
    Accuracy ±0.2 Vol. %
    Resolution 0.1 Vol. %
    Probe - Efficiency calculated
    Meas. Range 0 to 120%
    Resolution 0.1%
    Probe - CO2 calculated
    Meas. Range 0 to CO2 max
    Accuracy ±0.2 Vol. %
    Reaction time t90 Approx. 40 s

    Additional Details

    Industry Gas
    Type Flue Gas Analyser